چکیده:   (636 مشاهده)
Abstract
Introduction: HU is a standard numbering system in CT, which is proportional to x-ray attenuation and an indicator of relative tissue density. In CBCT, the x-ray 
attenuation degree is demonstrated with the grayscale. The present study aimed to 
determine the correlation between the grayscale in CBCT and HU (Hounsfield Unit) in MDCT.
Materials and Methods: This descriptive-analytical study was approved under the ethics code of IR.GUMS.REC.1398.451. A human dry mandible was immersed in a transparent cylindrical container to simulate soft tissue attenuation. The sample was scanned at three separate imaging centers using three CBCT units with the same brand. The scans were carried out once with standard irradiation conditions and once with high-resolution conditions. The mandible was scanned with a CT scan unit. The grayscale and HU of the enamel, dentin, cortical, and spongy bone were evaluated and compared with each other. Pearson’s correlation coefficient was used for data analysis at a significance level of 5% (P<0.05).
Results: In two standard and high-resolution conditions, there was a significant correlation (P<0.001) between the grayscale and HU. Given the positive values of the mean differences, it could be concluded that the gray level in CBCT has a positive correlation with HU in MDCT.
Conclusion: The grayscale in CBCT and HU in MDCT had a positive correlation, and according to this study, it is possible to calculate the HU from the gray level.
متن کامل [PDF 738 kb]   (386 دریافت)    
نوع مطالعه: پژوهشي | موضوع مقاله: عمومى
دریافت: 1401/8/17 | پذیرش: 1401/7/26 | انتشار: 1401/10/17

ارسال نظر درباره این مقاله : نام کاربری یا پست الکترونیک شما:
CAPTCHA

بازنشر اطلاعات
Creative Commons License این مقاله تحت شرایط Creative Commons Attribution-NonCommercial 4.0 International License قابل بازنشر است.

کلیه حقوق این وب سایت متعلق به مجله دانشکده دندانپزشکی دانشگاه علوم پزشکی گیلان می باشد.

طراحی و برنامه نویسی : یکتاوب افزار شرق

© 2024 CC BY-NC 4.0 | Journal of Dentomaxillofacial

Designed & Developed by : Yektaweb